Changes between Version 6 and Version 7 of WikiStart
- Timestamp:
- Mar 17, 2017, 11:01:45 AM (7 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
WikiStart
v6 v7 35 35 36 36 For each tension or current range, we take 8 samples between 0 and a value 37 close to the max of the range. For each sample we get the pic's output for 38 about 5 seconds, and we note the real value using the digital multimeter. 37 close to the max of the range. For each sample we get a few tousand values from pic's output, while we note the real value using the digital multimeter. 39 38 40 39 From this process, we found that that the AD623 output can't go down to 0, … … 45 44 we have a +/-7 unit error on the ADC's output. This is less than 0.18% of the 46 45 full range (or +/- 4.4mA on the 2.45A range input). 46 47 [[Image(0mA.gif, 50%, title="ADC values for 0mA",longdesc="ADC values for 0mA")]]